Home >> Products >> Thermal Solutions >> Fan performance / AMCA 210 Wind Tunnel >> LW-9293
LW-9293 Automatic Air Flow Rate and Pressure Measurement Apparatus
Based on following standards:
• ISO 5801-2007
• AMCA 210-07
• ASHRAE 51-2007
• IEC 61591-2005
• GB/T 1236-2000
Features
1. Follow AMCA 210 standard Fig.12 & 15 two-in-one structures
2. Automatic system control and data acquisition by PC
3. Nozzle's flow rate report prior to delivery is issued by Long Win,
which can be traced to an accredited lab for long-term traceability.
Experimental Items
1. Fan or fan tray's PQ performance
2. System Impedance Curve (SRC)
3. A standard flow rate generator
Specifications
1. Air flow rate: 230~30100 CFM
2. Accuracy of air flow rate: <3.5% INFS
3. Repeatability error: < 2 %
4. Static pressure: 0~150 mmAq
5. Overall dimension: 3.5 (W) × 12 (L) × 2.6 (H) m (Ref.)
6. Operation space: 5.5 (W) × 16 (L) × 2.8 (H) m (Ref.)
7. Power source: AC220V, 105 Amp, 3 phases or AC380V, 60 Amp, 3 phases
Technical Information
Download Catalogue