Home >> Products >> Fluid Mechanics >> Wind Tunnel >> LW-9069
LW-9069 Wind Tunnel for Hot Wire Anemometer Calibration
Specifications
1. Open-circuit suction-type structure
2. Test section : 100 × 100 × 400 mm
3. Flow field quality : on the same cross section
Velocity uniformity >99.5%
Turbulence intensity <0.5%
While more than 2 m/sec at 80% test section area from the center
4. Velocity : 0.2 ~ 40 m/sec
5. Power source : AC 220 V, 3 phase, 10 Amp
6. Dimension : 0.6 (W) × 2.2 (L) × 1.5 (H) m
1. Open-circuit suction-type structure
2. Test section : 100 × 100 × 400 mm
3. Flow field quality : on the same cross section
Velocity uniformity >99.5%
Turbulence intensity <0.5%
While more than 2 m/sec at 80% test section area from the center
4. Velocity : 0.2 ~ 40 m/sec
5. Power source : AC 220 V, 3 phase, 10 Amp
6. Dimension : 0.6 (W) × 2.2 (L) × 1.5 (H) m
Also compatible with 2-component balance unit
The test section and installation of hot wire anemometer